You are here: HomeScience OverviewLight Microscopy › New Ways to Super-Resolution: Reversible Chemical Reactions Control Fluorescent States

New Ways to Super-Resolution: Reversible Chemical Reactions Control Fluorescent States

Aug. 12, 2011
Fig. 1: Diffraction barrier limits the resolution in light microscopy.(a) Two objects in a distance above resolution limit (dmin) are easily resolvable.(b) Superposition of two fluorescence signals at a distance of around the resolution limit already shows an overlap but the origin of the signal is still distinguishable.(c) Two fluorescence signals overlap completely below the resolution limit and blur the origin of the signals (c).
Fig. 1: Diffraction barrier limits the resolution in light microscopy.(a) Two objects in a distance ... more
Fig. 1: Diffraction barrier limits the resolution in light microscopy.(a) Two objects in a distance ... Fig. 2: (a) To resolve objects below the resolution limit by localization microscopy based ... Fig. 3: (a) STED microscopy is based on the activation of fluorophores in a confocal volume ... Fig. 4: (a) For CHIRON blinking of the probe is induced by binding of copper(II) ions to a ... Fig. 5: (a) A micrograph can be reconstructed from movies recording the random blinking of the ... 

Recently, light microscopy has been revolutionized by novel approaches that circumvent the diffraction barrier the resolution limit of optical microscopes. Most of these novel methods are based on light-controlled switching of the labels fluorescent states and therefore require use of additional or more intense laser excitation lines. To overcome these demands a new probe has been developed which is controlled by a reversible chemical reaction thereby reducing the demands on the microscopes.

Limitations of Light Microscopy

In 1873 Ernst Abbe postulated one of the fundamental limitations of microscopy [1]. He suggested that it is impossible to resolve two objects within a distance of about half of the wavelength used. The so called diffraction barrier results in a resolution limit of roughly 200 nm for visible light (fig. 1). In the past decade different methods have been developed to bypass this limitation.

Most prominent among these methods are the Stimulated Emission Depletion (STED) microscopy [2] and the single-molecule based localization microscopy approaches (Stochastic Optical Reconstruction Microscopy - STORM, Photo-Activation and Localization Microscopy - PALM) and their variations [3,4,6]. Localization microscopy methods are based on the precise localization single fluorophores by use of mathematical fitting algorithms to an accuracy of about 1 nm. By driving most fluorophores of the observed sample into a non-fluorescent state it is possible to localize the central point of a few remaining fluorescent spots - and the position of the fluorophore respectively - with high accuracy [5]. By repeating this procedure several times a high resolution image can be obtained (fig. 2).

Microscopy to Nanoscopy - Switching is Key

While all localization techniques share similar mathematical algorithms to localize single fluorophores their main difference is the method used to switch the fluorophores to a dark state.



Ima
ging & Microscopy Issue 4 as free epaper or pdf download

 

In PALM a few photo-activatable proteins are statistically activated by use of more energetic light and afterwards those activated fluorescent proteins are imaged till they are photo-bleached with an imaging laser [4]. In STORM all fluorophores are in a bright state before they are deactivated by a laser pulse with high power. Reactivation of the fluorophores is driven either by a blue shifted laser pulse with low power to activate only a small subpopulation or by use of reagents like mercaptoethylamine (MEA) [3,7].

Obviously, the use of high power lasers or of laser lines with different energy leads to the necessity of special microscopes. To generalize localization microscopy methods it is possible to say that the fundamental underlying principle is the cycling of probes between a fluorescent (on) and a non-fluorescent (off) state (fig. 3). Additionally the ratio of activated fluorophores to those in an off state strongly depends on the density of fluorophores in the observed sample and a value of 1:200 has been suggested as a good parameter for most biological samples [8]. The ultimate probe would therefore be based on a random switching process which is light-independent but can be controlled by some other mechanism (fig. 3c).

Chemistry Provides Alternative Switching Mechanisms

Uncoupling of the switching process from light has many promising advantages. Due to the high laser power needed for photo-switching the introduction of alternative mechanisms lowers the radiation and therefore the photo toxicity for biological samples. Another advantage is the minimization of requirements for the experimental setup. To realize the idea of light-independent switching processes an ion sensor has been modified. The requirements of a probe suitable for localization microscopy are: (i) the on/off ratio has to be adjustable (ii) the probe has to show bright on and dark off states (iii) the on times of the probe have to be around 100 ms and (iv) the probe has to be labeled to some biological target. A probe which has been developed a few years ago to observe chemical reactions in thermodynamic equilibrium at single-molecule level fulfilled all these requirements [9].

The essential parts of the probe are on the one hand a fluorophore and on the other hand a ligand which is able to bind free ions. The chelating ligand used is a bipyridine which efficiently binds copper(II) ions. Due to the spatial proximity of the copper-bipyridine-complex fluorescence of the dye is quenched (fig.4a). By increasing the concentration of copper(II) ions in the measurement buffer a strong influence on the on-duration could be observed while the off-duration of the fluorophores remained unaltered. This can be explained by the fact that with increasing ion concentration the probability to bind an ion to the ligand increases while the energy of the bond is unaffected. Due to the nature of the probe, this novel technique was named chemically improved resolution for optical nanoscopy (CHIRON) [10]. The probe design is based on a double stranded DNA containing a biotinylated end allowing attachment of the probe to streptavidin tagged targets (fig. 4b) to fulfill the requirement for biological applicability.

Related Articles :

Keywords: CHIRON Dirk-Peter Herten dSTORM Fluorescence Microscopy Light Microscopy nanoscopy PALM Photo-Activation and Localization Microscopy STED Stimulated Emission Depletion Microscopy Stochastic Optical Reconstruction Microscopy Super-Resolution Imaging Super-Resolution Microscopy

Email requestCompany Homepage

University Heidelberg
Im Neuenheimer Feld 267
69120 Heidelberg
Germany

Web: http://www.uni-heidelberg.de/

RSS Newsletter