Laser-Guided Atomic Force Microscopy

Precision Photonics Meets Nanotechnology

Aug. 01, 2011
Fig. 1: Artist’s rendition of ultrastable AFM. The typical force-sensing laser (gold) is complimented by two additional lasers (red & green) which scatter off the tip and a fiducial mark patterned onto the sample, respectively. The resulting signals are used to measure and thereby control the position.
Fig. 1: Artist’s rendition of ultrastable AFM. The typical force-sensing laser (gold) is ... more
Fig. 1: Artist’s rendition of ultrastable AFM. The typical force-sensing laser (gold) is ... Fig. 2: Simplified optical schematic of the instrument. Two stabilized lasers are focused by a ... Fig. 3: Demonstration of instrument stability. (a) Initially, the tip position was measured in x ... Fig. 4: Demonstration of a tripod for AFM. (a) A photograph taken in low light with a fast shutter ... Fig. 5: Optically locating a patch of membrane proteins for subsequent AFM study. (a) A large area ... 

Application of Atomic Force Microscopy (AFM) to a broad range of research, and biological systems in particular, is hampered by two longstanding technical problems: mechanical drift and finding sparsely distributed samples. By adapting ideas from the optical-trapping community, we have made significant progress in addressing both of these issues, including a 100-fold improvement in the stability of AFM at ambient conditions.

The atomic force microscope (AFM) is an invaluable instrument in nanoscience research and development. AFM's broad use arises from its ability to probe and manipulate a wide array of materials in an equally wide variety of operating conditions. For instance, scientists have used AFMs to study nanometer-scale structures and dynamics in vacuum over five decades in temperature (0.01-1,000 K), to image and to mechanically unfold individual proteins at physiologically relevant conditions, and even to visualize soil and ice grains on another planet (NASA Phoenix Mars Lander).

Two long-standing problems in AFM are mechanical drift and finding sparsely distributed samples. Drift, particularly at room temperature, limits observation time because it is difficult to distinguish from the true signal. Additionally, it is time consuming to find desirable regions for detailed study because the samples are often randomly absorbed onto a substrate. This process can also degrade the mechanical and chemical integrity of the tip.

An Optically Based Reference Frame for AFM

We solved this pair of problems by integrating two additional highly stabilized lasers to supplement traditional force detection. Initially, the motivation was to detect and thereby control the three dimensional (3D) position of the tip and the substrate via scattered light, a strategy inspired by precision optical-trapping techniques [1-3]. In an artist's rendition of the central components (fig. 1), the tip-detection laser beam is red. A second beam (green) detects the position of the sample via scattering off a fiducial mark on the substrate. These two measurements are independent of and complementary to the traditional force measurement, based on cantilever deflection (gold).



Ima
ging & Microscopy Issue 4 as free epaper or pdf download

 

These advances rely on detecting back-scattered light from a focused laser beam (fig. 2). Using a focused laser beam enables 3D determination of the object's location, in contrast to the 1D information derived in a traditional interferometer. Sub-Ångstrom sensitivity was achieved by minimizing a variety of optical noise sources.
We solved the first problem, drift, by referencing the position of the tip to the position of the sample via an optically based reference frame. The resulting optically stabilized AFM drifts by only a few Ångstroms over an hour operating at "real-world" conditions (in air at room temperature) [4]. This lateral stability arises because it is not the absolute stability of the lasers that matters but rather their differential-pointing stability.

We solved the second problem, locating regions of interest, by enhancing an existing label-free optical-imaging technique [5, 6]. In doing so, we employed the same laser used for tip stabilization. The resulting optical images of thin, transparent biological structures were registered to subsequent AFM images with nanometer-scale precision (<40 nm) because the tip is aligned to the optical axis of the imaging laser.

Long-Term Stability at Real-World Conditions

Unwanted mechanical drift between the tip and the sample places fundamental limits on AFM instrument performance and decreases utility. Drift can be minimized by operating in vacuum or under cryogenic conditions, but this strategy precludes many important investigations that require operating in air or fluid at room temperature such as imaging and probing the dynamics of biological assemblies. If drift were absent from an AFM, it would be possible to (i) enhance the imaging signal by scanning slowly to average the Brownian motion of the cantilever, (ii) return the tip to a precise feature in an image (e.g., a protein domain), (iii) hover the tip over a feature to study local dynamics (e.g., conformational fluctuations), and (iv) control the 3D position of the tip when disengaged from the surface (e.g., force spectroscopy).

Our optically based, active feedback virtually eliminates drift at room temperature. We measure environmentally induced noise and then use active feedback via a pair of three-axis piezo-electric stages to mitigate its effect. Tip-sample stability is achieved because there is essentially no differential-pointing noise between these added lasers (0.2 Å, Δf = 0.1-50 Hz) [7]. Thus, a local, optically based reference frame minimizes the traditional need for a stiff "mechanical loop", the physical path through the frame of the instrument that connects the AFM tip to the substrate.
The benefit of this stabilization method is immediately apparent from fig. 3. In the top panel, the onset of feedback is shown to essentially eliminate a substantial drift. Moreover, as shown in the bottom panel, feedback provides a means for arbitrary control of the tip position in two (or three) dimensions with sub-Ångstrom precision.